MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1550400-38
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Angina (1710); Thrombosis (2100)
Date of Event 11/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device was not returned for evaluation.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this event.The reported patient effects of thrombosis and angina are listed in the xience sierra everolimus eluting coronary stent systems instructions for use as known patient effects of coronary stenting procedures.The investigation was unable to determine a conclusive cause for the reported difficulty to deploy (wall apposition).A conclusive cause for the reported patient effects and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The 4.0x23mm xience sierra is filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that on (b)(6) 2018, the patient presented with angina.The procedure was to treat a lesion located in a moderately tortuous, moderately calcified concentric mid to distal right coronary artery (rca) a 4.0x23mm xience sierra stent and a 4.0x38mm xience sierra stent were deployed in an overlapped manner.Timi flow changed from 2 to 3.Intravascular ultrasound (ivus) was confirmed to successfully complete the procedure.It was confirmed that the patient was compliant with dual antiplatelet drug therapy after the procedure.On (b)(6) 2018, the patient was re-hospitalized due to chest pain.In-stent thrombus was confirmed via angiography.The stent was recoiled.Balloon angioplasty was performed with a 4.0x12mm unspecified balloon.On (b)(6) 2018, the patient was discharged from the hospital.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8086961
Report Number2024168-2018-08936
Device Sequence Number11192955
Product Code NIQ
UDI-Device Identifier08717648227547
UDI-Public08717648227547
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/25/2021
Device Catalogue Number1550400-38
Device Lot Number8031841
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2018
Initial Report FDA Received Date11/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age81 YR
-
-