| Catalog Number |
1003382-28 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was returned for analysis.A visual and dimensional inspection was performed.The reported stent dislodgement was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported stent dislodgement.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during removal of the protective sheath, without resistance, the 5.0x28 mm ultra stent dislodged inside the protective sheath.The device was not used.A same size ultra stent delivery system was used to complete the procedure.There was no reported patient involvement or a clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|