AV-TEMECULA-CT ACS MULTI-LINK RX ULTRA CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1003382-28
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was returned for analysis.A visual and dimensional inspection was performed.The reported stent dislodgement was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported stent dislodgement.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that during removal of the protective sheath, without resistance, the 5.0x28 mm ultra stent dislodged inside the protective sheath.The device was not used.A same size ultra stent delivery system was used to complete the procedure.There was no reported patient involvement or a clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACS MULTI-LINK RX ULTRA CORONARY STENT SYSTEM
Common Device Name
CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8087009
Report Number2024168-2018-08937
Device Sequence Number1599045
Product Code MAF
UDI-Device Identifier08717648035326
UDI-Public08717648035326
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
P020047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Catalogue Number1003382-28
Device Lot Number7110641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2018
Initial Report FDA Received Date11/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-