H10: manufacturing review: a dhr review was not required as this lot met all release criteria.There was nothing found to indicate a manufacturing-related cause for this event.Investigation summary: one seeker was returned for evaluation.The device was returned with a detached tip, therefore, the investigation is confirmed for tip detachment.The catheter measured approximately 149.0cm from the strain relief.Therefore, the length of detached segment was approximately 1 centimeter.The root cause could not be determined based upon available information.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H11:section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|