| Catalog Number |
8590855 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Procedure: lumbar decompression and bone graft fusion it was reported that intra-op, screw slipped during screw implantation in the surgery.No patient complications occurred as a result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: visual and macroscopic inspection confirmed the threads of the break off screw have been damaged.The thread crest and flank damage appears to have initiated at the start of the thread, and is consistent around the damaged portion of the thread.Functional evaluation with a sample mas found the set screw unable to be fully engaged in the mas head.The upper break off portion has been broken from the lower threaded section of the screw.The hex of the screw has also been damaged.The above observations are consistent with misalignment of the mas and set screw threads during construct assembly.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|