| Model Number |
TRS175SB2 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One trs175sb2 was received in opened original packaging, the reported catalog number was verified.Visual inspection found that the tongs had broken through the bag.Inspection also found that the draw string had bunched up and appeared to be caught in the handle.The condition of the draw string caused premature cinching of the bag and the tongs broke through.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A two-year review of complaint history revealed that there have been a total of 13 complaints, regarding 15 devices for this device family and failure mode.During the same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have benn found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following; only physicians trained in the techniques of laparoscopic or minimally invasive surgery and the use of retrieval bags to remove tissue should use the product.The anchor tissue retrieval system is for single patient use only and should not be re-sterilized.The ability to effectively clean and reuse this single use device has not been established and subsequent re-use may adversely affect the performance, safety and/or stability of the device.The use of morcellators within the anchor tissue retrieval system has not been evaluated.The anchor tissue retrieval system is not recommended for use with power morcellators.Care should be taken when using sharp and electrosurgical devices.Note the size of the trocar when removing a specimen through the trocar cannula and seal system.If required, enlarge the port site to aid in the removal of the anchor tissue retrieval system.
|
| |
|
Event or Problem Description
|
|
The customer reported that the metal prong had ripped through the bag, while trying to remove the specimen during a thoracic procedure on (b)(6) 2018.They had to use 2 bags, each bag was only deployed once, there was some resistance while releasing the bag.The bag was not pre-tested, and the surgical team was familiar with the product.There was no patient injury reported and there was a delay in surgery, however no time length was given.The surgery was successfully completed with a 3rd bag of the same type.This report is being raised based on device malfunction with potential for injury upon reoccurrence.
|
| |
|
Search Alerts/Recalls
|