9 days post primary, after a session of physiotherapy, the patient has felt severe pain and could not stand.X-ray revealed a periprosthetic fracture, so on the next day the patient underwent to a revision surgery: the femur was cabled with 2 cerclage wires and the stem was converted to quadra stem.The surgery was completed successfully.It was not believed that the fracture had been due to aggressive physiotherapy, but they believed that it could have been due to rapid recovery of the patient coupled with an initial valgus position and undersized stem - that might have subsided due to very early mobilization into neutral/varus position (and then became "oversized" have contributed to the fracture).
|