Investigation summary: customer returned (12) 1cc, 8mm, 31g syringes (2 loose, 10 in a sealed poly bag) from lot # 7241532.Customer states that they are having a hard time pressing the plunger and there are burrs on the needles.All returned syringes were tested and all were able to draw and expel properly without any observed defects.All returned syringes were also tested for point geometry, lube, and cannula od (specs: outer diameter for 31g: 0.0100¿-0.0105¿).All observations fall within specifications.A review of the device history record was completed for batch# 7241532.All inspections were performed per the applicable operations qc specifications.There were three (3) notifications [200712892, 200712495, 200712933] noted that did not pertain to the complaint.Based on the samples / photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure.Root cause cannot be determined at this time as the issue is unconfirmed.
|