Model Number 945ECLC
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis found that the device had a connection failure due to twisted pins on the stimulator connector.The device was disassembled, cleaned, replaced connector pcba, applied loctite to connectors, reassembled, and placed into the burn-in chamber to check for any heat related failures.The device was removed and tested in accordance with manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care provider (hcp) reported that the device was not producing a high stimulus; when both stimulators were plugged in they were getting an electrical stim failure; tried with two other stimulators on the same device and had the same issue.There was no known patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM-ECLIPSE® CONTROLLER
Common Device Name
STIMULATOR, ELECTRICAL, EVOKED RESPONSE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key8090442
Report Number1045254-2018-00626
Device Sequence Number1684480
Product Code GWF
UDI-Device Identifier00643169071872
UDI-Public00643169071872
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K050798
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number945ECLC
Device Catalogue Number945ECLC
Device Lot Number208362595
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2018
Initial Report FDA Received Date11/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-