| Model Number |
279722890 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint : (b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of instrument (b)(4) broke inside the screw head.The bone was very hard even though the surgeon tapped the whole before insertion.The screw was removed and replaced with another one.The tip of instrument (b)(4) got stripped during final tightening.Patient consequence? :no.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual examination at the macroscopic level found that the fracture was located at the driver¿s distal tip.A fracture analysis report was conducted on the returned device.It was noted plastic deformation at the hexlobes and torsional shear markings following a circular pattern.This plastic deformation at the hexlobes indicates a torsional overload.This suggests that the driver underwent a quasi-static overload torsional shear failure.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.A definitive root cause for the quick-connect driver distal tip fracturing cannot be positively determined.However, as noted in the fracture analysis report it is suggested that the driver underwent a quasi-static overload torsional shear failure.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|