BOSTON SCIENTIFIC CORPORATION MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7575
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Event date was not reported and approximated (b)(6) 2018.
 
Event or Problem Description
Reportable based on device analysis completed on 24-oct-2018.A 2.50mm x 9mm maverick 2 balloon catheter was returned with no reported allegation, however, investigation results revealed the shaft detached/separated.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube.The hypotube is separated 66.1cm from the hub.Microscopic examination revealed no additional damages.There is blood present in the guide wire lumen and on the balloon folds.The balloon is partially loosely folded.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis found multiple kinks and a separated hypotube.The hypotube appeared to be kinked prior to separation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8090978
Report Number2134265-2018-62871
Device Sequence Number11114275
Product Code LOX
UDI-Device Identifier08714729369899
UDI-Public08714729369899
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2020
Device Model Number7575
Device Catalogue Number7575
Device Lot Number0020958260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2018
Initial Report FDA Received Date11/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-