Model Number CQ7554
Medical Device Problem Codes Leak/Splash (1354); Peeled/Delaminated (1454)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Expiry date 01/2021.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a forearm fistula, contrast was allegedly leaking at the distal end of the pta balloon upon inflation.Reportedly, another balloon was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.A visual inspection found peeled pebax on the proximal cone of the balloon.Therefore, the investigation is confirmed for peeled pebax.The device was inflated with water and no leaks were noted to the balloon or remained of the device.Therefore, the investigation is unconfirmed for the reported leak.The definitive root cause for the identified peeling pebax or reported leak could not be determined based upon the available information.An investigation is currently open to address the peeling pebax issue.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiry date 01/2021), section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a forearm fistula, contrast was allegedly leaking at the distal end of the pta balloon upon inflation.Reportedly, another balloon was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8091044
Report Number2020394-2018-02099
Device Sequence Number13874016
Product Code DQY
UDI-Device Identifier00801741062834
UDI-Public(01)00801741062834
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQ7554
Device Catalogue NumberCQ7554
Device Lot NumberRECN3123
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2018
Initial Date Received by Manufacturer 10/29/2018
Supplement Date Received by Manufacturer12/15/2018
Initial Report FDA Received Date11/20/2018
Supplement Report FDA Received Date01/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-