• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number A2025-080
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The product was not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation was unable to determine a conclusive cause for the reported balloon rupture.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat an unspecified peripheral artery.A 2.5 x 80 mm armada 14 balloon catheter was used; however, the balloon ruptured at 6 atmospheres.Therefore, the device was replaced with another unspecified armada balloon catheter to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8091069
Report Number2024168-2018-08964
Device Sequence Number1747536
Product Code LIT
UDI-Device Identifier08717648156038
UDI-Public08717648156038
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Catalogue NumberA2025-080
Device Lot Number8041941
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2018
Initial Report FDA Received Date11/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-