MAUDE Adverse Event Report: AV-TEMECULA-CT ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE WITH HYDROCOAT HYDROPHILIC COATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE WITH HYDROCOAT HYDROPHILIC COATING Back to Search Results
Catalog Number 1001782S-HC
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/13/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the guide wire during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the mildly tortuous, concentric anatomy and/or other devices resulted in the reported tip detachment.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was performed to treat a mildly tortuous and concentric lesion in the mid iliac artery.A 014 bmw hydro guide wire was advanced to the lesion and an intravascular ultrasound catheter was advanced over the guide wire without resistance; however, the tip of the guide wire separated.A snare was used to remove the separated tip.The procedure was successfully completed with a new 014 bmw hydro guide wire.There were no adverse patient sequela and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE WITH HYDROCOAT HYDROPHILIC COATING
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8091166
Report Number2024168-2018-08965
Device Sequence Number1279600
Product Code DQX
UDI-Device Identifier08717648025990
UDI-Public08717648025990
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K013833
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2020
Device Catalogue Number1001782S-HC
Device Lot Number8030271
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2018
Initial Report FDA Received Date11/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient Weight95
-
-