The associated complaint device was not returned.Without the actual product involved and/or device information, our investigation cannot proceed.The medical investigation concluded that the undated images provided revealed progression of the laterally positioned distal femur along with a varus deformity.Currently, we are unable to rule out initial lateral alignment of the distal femur and the lack of pre-operative adaptations to prevent movement of limb segments as contributing factors to the reported event.The patient impact beyond aesthetics, altered joint wear/gait, possible revision surgery, additional rehabilitative therapy, and radiological exposure cannot be concluded but if there is correction of this deformity, another procedure would be required or a permanent deformity may impact their function.No further medical assessment can be rendered at this time.If the device or new information is received in the future, the complaint can be re-opened.No further actions are being taken at this time.
|