Model Number I20-13/C70F SA
Medical Device Problem Codes Loss of or Failure to Bond (1068); Leak/Splash (1354)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/26/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.If additional information is obtained at a later time that is pertinent to this event, a follow-up report will then be submitted.
 
Event or Problem Description
The patient was initially treated with the afx abdominal aortic aneurysm stent.Approximately one (1) year post initial procedure, a type 1b endoleak was detected ct scan.There have been no additional patient sequelae reported.
 
Additional Manufacturer Narrative
At the completion of the clinical evaluation and based on the information received, there was substantial evidence to support the reported event of type ib endoleak.The complaint is most likely user-related (off-label absence of aortic aneurysm and incomplete covering of left iliac aneurysm).No contributing factors were identified; unable to determine iliac measurements pre index procedure due to lack of documentation.Procedure related harms for this complaint could not be determined.The final patient status was not reported.The manufacturing lot review confirmed all devices met specifications prior to release.These types of events will be monitored and trended as part of the quality system.Conclusion code - remove 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
LIMB STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key8091661
Report Number2031527-2018-00875
Device Sequence Number16518546
Product Code MIH
UDI-Device Identifier00818009011442
UDI-Public(01)00818009011442(17)170808
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2017
Device Model NumberI20-13/C70F SA
Device Lot Number1253169-012
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/23/2018
Supplement Date Received by Manufacturer10/23/2018
Initial Report FDA Received Date11/20/2018
Supplement Report FDA Received Date01/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX-BIFURCATED STENT GRAFT:1343696-004; AFX-LIMB STENT GRAFT:1257552-002
Outcome Attributed to Adverse Event Other;