| Catalog Number |
07K76-30 |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
All available patient information was included.Additional patient details are not available.An evaluation is in process.A final report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer observed a falsely elevated prolactin result on the architect i2000sr analyzer.The following data was provided: initial greater than 200, repeated 326, 14.40 ng/ml.Progesterone results of 16.04 and 12.30 ng/ml were also provided.The patient was on the following medications: levothyroxine, cyclobenzaprine, domperidone, glucosamine.There was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Catalog number, size, and code was updated from 07k76-35 to 07k76-30.Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and accuracy testing.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.The product was not available for return.An internal panel was tested with retained kits of the likely cause reagent lot and accuracy testing met all specifications.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
|
| |
|