BARD ACCESS SYSTEMS SITE-RITE VISION II 20MM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770169
Device Problem No Display/Image (1183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.Results are expected soon.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed: the ultrasound scans are too dark.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample(if available), applicable fmea documents, labeling and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of ultrasound scans are too dark was unconfirmed.During inspection the probe was tested and worked normally, no root cause confirmation since the issue could not be reproduced.The device was discarded due to being discontinued product.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed: the ultrasound scans are too dark.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE VISION II 20MM
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8092453
MDR Text Key128078515
Report Number3006260740-2018-03333
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00801741125348
UDI-Public(01)00801741125348
Combination Product (y/n)N
PMA/PMN Number
K132942
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770169
Device Catalogue Number9770169
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2018
Event Location Hospital
Date Manufacturer Received08/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-