BARD ACCESS SYSTEMS SITE~RITE 6 PROBE; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770001
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.Results are expected soon.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Per biomed: the probe is casting shadows in the image.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of probe is casting shadows in the image is confirmed, as the probe has lens and keypad damage and is use related.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Per biomed: the probe is casting shadows in the image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 6 PROBE
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8092456
Report Number3006260740-2018-03330
Device Sequence Number2178325
Product Code ITX
UDI-Device Identifier00801741125744
UDI-Public(01)00801741125744
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K071204
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9770001
Device Catalogue Number9770001
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2018
Event Location Hospital
Initial Date Received by Manufacturer 11/02/2018
Supplement Date Received by Manufacturer06/26/2019
Initial Report FDA Received Date11/20/2018
Supplement Report FDA Received Date07/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-