Model Number HGK0008-20
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code Blood Loss (2597)
Date of Event 10/23/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is not accessible for testing as it remained implanted in patient.A review of the complaint device history records indicated that the graft was processed and inspected according to established procedures and was therefore released following acceptable quality inspections and tests.Specifically, no anomaly was evidenced in the collagen coating records.The review of the water permeability testing records of products coated on the same day as the complaint device indicated values well within product specifications.One retention sample coated on the same day and under the same conditions as the complaint device underwent water permeability testing.The test result indicated a value well within product specifications (< 5 ml/cm²/min).The review of post-marketing historical data indicated that no other similar complaint was reported for the same product family and same lot number.No conclusion can be drawn.However, the conducted investigation and testing performed would tend to indicate that the product was not defective at the time of manufacturing.Please note that, as per the product instructions for use, the use of knitted vascular grafts as a conduit for cannulation is not an approved indication.Indeed, hemagard knitted vascular grafts are indicated for surgical repair, bypass, or replacement of arteries in the treatment of aneurysmal and occlusive disease of the abdominal aorta, visceral arteries, and peripheral arteries.
 
Event or Problem Description
The concerned graft was used as an extracorporeal vessel channel.The graft leaked during surgery.Patient endangered but not injured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMAGARD KNITTED
Common Device Name
VASCULAR POLYESTER GRAFT
Manufacturer (Section D)
INTERVASCULAR SAS
z.i. athelia i
la ciotat cedex, 13705
FR  13705
MDR Report Key8093977
Report Number1640201-2018-00023
Device Sequence Number11259465
Product Code DSY
UDI-Device Identifier00384401014133
UDI-Public00384401014133
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K964625
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date12/31/2021
Device Model NumberHGK0008-20
Device Catalogue NumberHGK0008-20
Device Lot Number17A05
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2018
Initial Report FDA Received Date11/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age80 YR
Patient Weight60
-
-