• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. SILK SUTURE 18"(45CM) 6-0 BLK; SUTURE, NONABSORBABLE, SILK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SILK SUTURE 18"(45CM) 6-0 BLK; SUTURE, NONABSORBABLE, SILK Back to Search Results
Catalog Number 706G
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that a patient underwent a ptosis procedure on (b)(6) 2018 and suture was used.During use, the needle popped off without any tension.Case completed with another device of the same product code.There were no adverse patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent to the fda: 12/17/2018.Investigation summary: it was reported performance pull off suture needle.A empty folder, a dispensed needle suture and a detached needle were received for analysis.This product code is double armed.During the visual inspection of detached needle, the swage and attachment area were as expected.The barrel hole was examined under 20x magnification and remnant suture was noted.The suture end is severely cut (damaged), resulting in a clip off defect.In the inspection of the needle/suture combination, the swage and attachment area were as expected.The manufacturing records couldn't be reviewed as the batch number is unknown.Per the sample condition the assignable cause of the needle pull off, is a clip off defect, this defect is caused by excessive pressure on the suture during swaging of the needle and consequently breaking.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILK SUTURE 18"(45CM) 6-0 BLK
Common Device Name
SUTURE, NONABSORBABLE, SILK
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key8094220
Report Number2210968-2018-77283
Device Sequence Number16830553
Product Code GAP
UDI-Device Identifier10705031008724
UDI-Public10705031008724
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number706G
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2018
Initial Date Received by Manufacturer 10/26/2018
Supplement Date Received by Manufacturer11/21/2018
Initial Report FDA Received Date11/21/2018
Supplement Report FDA Received Date12/17/2018
Patient Sequence Number1
-
-