• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910  12/14 ARTICUL 44MM M SPEC+1.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910  12/14 ARTICUL 44MM M SPEC+1.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1365-61-000
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Adhesion(s) (1695); Pain (1994); Discomfort (2330); Hypoesthesia (2352); No Code Available (3191)
Date of Event 10/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient's metal on metal hip construct was revised due to pain.Surgeon removed the metal liner and replaced it with a poly liner.Doi: (b)(6) 2009, dor: oct 29, 2018, left hip.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate. .
 
Additional Manufacturer Narrative
Product complaint # (b)(4).
 
Event or Problem Description
After review of medical records, the patient had complaints of discomfort, numbers and elevated metal ions.Then the patient was revised for failed metal on metal left tha.Operative notes reported that patient had fluid in his hip and extensive scars were removed.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: corrected: h6 (device).Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot :null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12/14 ARTICUL 44MM M SPEC+1.5
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8094255
Report Number1818910-2018-76134
Device Sequence Number1850645
Product Code JDI
UDI-Device Identifier10603295034032
UDI-Public10603295034032
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K060031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1365-61-000
Device Catalogue Number136561000
Device Lot Number2467254
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/29/2018
Supplement Date Received by Manufacturer12/03/2018
01/05/2020
02/13/2020
Initial Report FDA Received Date11/21/2018
Supplement Report FDA Received Date12/04/2018
01/07/2020
02/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-