• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. IPC® HANDPIECE - XPS® STRAIGHTSHOT® M4; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. IPC® HANDPIECE - XPS® STRAIGHTSHOT® M4; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE Back to Search Results
Model Number 1898200TRF
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis results were not available as of the date of this report.  a follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care provider (hcp) reported that the handpiece overheated suddenly during a functional endoscopic sinus surgery (fess) with nasal polypectomy procedure.Within minutes, the device was hot to touch, then the handpiece stopped working.The case was completed but there was an hour delay as they had to get the other device from another hospital by cab.There was no patient involvement.
 
Additional Manufacturer Narrative
Analysis found that the reported complaint of excessive heat could not be verified because the handpiece was not recognized by the console.The device had motor failure and the cable failed earth resistance test.The housing of the device was disassembled and found that the seals, bearings and motor were corroded due to dried saline.All seals, bearings, motor/cable were replaced.The device was repaired, cleaned and tested to manufacturing specifications.(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPC® HANDPIECE - XPS® STRAIGHTSHOT® M4
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key8094551
Report Number1045254-2018-00639
Device Sequence Number2025882
Product Code ERL
UDI-Device Identifier00613994598301
UDI-Public00613994598301
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K041413
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1898200TRF
Device Catalogue Number1898200TRF
Device Lot Number207884795
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2018
Supplement Date Received by Manufacturer11/27/2018
Initial Report FDA Received Date11/21/2018
Supplement Report FDA Received Date12/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-