| Model Number |
CD1215-36Q |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient presented with a device that exhibited possible premature battery depletion.No further information is available.
|
| |
|
Event or Problem Description
|
|
New information received on nov 29, 2018, indicates that no resolution was reported.
|
| |
|
Event or Problem Description
|
|
New information received on (b)(4) 2019, indicates that the patient presented for an explant procedure on (b)(6) 2019.The patient was stable before, during, and after the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was normal.No anomalies were found.It was found that the longevity estimate given to the field by the programmer was incorrect.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number:2017865-2019-09425.
|
| |
|