• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS HUBER PLUS 22G X 0.75"; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS HUBER PLUS 22G X 0.75"; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of recr1129 showed thirteen other similar product complaint(s) from this lot number.The complaints for this lot number (recr1129) have been reported from the same facility in (b)(6).
 
Event or Problem Description
It was reported the device package was deformed on 14 devices (the sterility of device was suspected to be not in good condition).Devices were not used on patients.This report addresses the first device.
 
Event or Problem Description
It was reported the device package was deformed on 14 devices (the sterility of device was suspected to be not in good condition).Devices were not used on patients.This report addresses the first device.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of deformed packaging is confirmed and the cause appeared to be associated with prolonged exposure to excessive heat.Fourteen 22 ga x 0.75 in huber plus infusion sets in unopened packaging were returned for investigation.The plastic was observed to be warped and the flanges were partially separated from the tyvek packaging.The presence of the seal was observed; however, a complete separation from the seal was observed on two of the fourteen samples due to the deformation of the trays.This type of damage can occur from prolonged exposure to excessive heat, and the damage was most likely associated with shipping or environmental conditions outside bard control.The product ifu found on www.Bardaccess.Com states ¿do not use if package is damaged¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUBER PLUS 22G X 0.75"
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8095592
Report Number3006260740-2018-03340
Device Sequence Number19295423
Product Code FPA
UDI-Device Identifier00801741065712
UDI-Public(01)00801741065712
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K993848
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number012234
Device Lot NumberRECR1129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2018
Event Location Hospital
Initial Date Received by Manufacturer 10/29/2018
Supplement Date Received by Manufacturer12/19/2018
Initial Report FDA Received Date11/21/2018
Supplement Report FDA Received Date01/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-