| Model Number |
N85 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device handheld monitor was received for evaluation.The battery had a date code of (b)(6) 2016, which is past the two-year limit where it would normally be replaced at factory service.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the unit was not really used on a patient.There was a code blue on a patient, they were going to use this unit on the patient to check spo2 and co2.But when the unit was turned on, it was red and did not give any kind of readings at all.
|
| |
|