• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problems Overheating of Device (1437); Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/26/2018
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of low power was not confirmed.Product failed functional testing with blade stall error and overheating.Cause of overheating and errors is a corroded motor/gearbox.The motor/gearbox assembly could not be removed from the housing for further assessment due to corrosion.The complaint of low power was not confirmed.The root cause of overheating has been determined to be corrosion of the motor and gearbox assembly.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Event Description
It was reported that device has low power.The event occurred before the procedure.During the investigation the device overheated.No injuries were reported.
 
Manufacturer Narrative
Additional information received by the manufacturer has identified that this event should be re-evaluated for mdr reporting.The reassessment determined that the issue does not meet the threshold for reporting and is a non-reportable event because this is a duplicate of the event reported on complaint (b)(4).This report was made based upon information which smith & nephew had not been able to investigate or verify prior to the required reporting date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key8095996
MDR Text Key128195735
Report Number1643264-2018-00904
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received11/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-