• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER COOL-TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER COOL-TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number ACT2020
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during ablation, the device detects the needle and shows that the impedance is too high.It was replaced by the new device but still detects the needle had an impedance that is too high.The procedure was completed by using another device.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: evaluation summary: one device was received for evaluation.The investigation found the device to function normally and within specifications.The water circulated normally through the device.The device read the temperature properly.No error message was noted.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOL-TIP
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key8096465
Report Number1717344-2018-01528
Device Sequence Number1903003
Product Code GEI
UDI-Device Identifier10884524001210
UDI-Public10884524001210
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K973297
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/15/2020
Device Model NumberACT2020
Device Catalogue NumberACT2020
Device Lot Number73050165X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2018
Initial Date Received by Manufacturer 10/29/2018
Supplement Date Received by Manufacturer12/20/2018
Initial Report FDA Received Date11/21/2018
Supplement Report FDA Received Date01/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-