| Model Number |
9004 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Reocclusion (1985)
|
| Date of Event |
06/14/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.A lot history review revealed 1 additional reocclusion complaint associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual samples were not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was possibly related to the study devices, but not related to the procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported through a post market clinical study that during the index procedure, two lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the right superficial femoral artery (sfa).Approximately 1 month after the index procedure, the patient¿s right sfa was reportedly reoccluded.No revascularization has been performed at this time.The investigator assessed that the event was possibly related to the study devices, but not related to the procedure.The samples were discarded by the user facility and are not available for evaluation.No adverse patient effects were reported.This is one of two products involved with the reported event and the associated manufacturer¿s report number is 3006513822-2018-00274.
|
| |
|
Additional Manufacturer Narrative
|
|
The operator of device - other had the following information added "cardiovascular internal medicine." the complainant occupation - other had the following information added "cardiovascular internal medicine." analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.A lot history review revealed 1 additional reocclusion complaint associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual samples were not returned for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was possibly related to the study devices, but not related to the procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported through a post market clinical study that during the index procedure, two lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the right superficial femoral artery (sfa).Approximately 1 month after the index procedure, the patient¿s right sfa was reportedly reoccluded.No revascularization has been performed at this time.The investigator assessed that the event was possibly related to the study devices, but not related to the procedure.The samples were discarded by the user facility and are not available for evaluation.No adverse patient effects were reported.This is one of two products involved with the reported event and the associated manufacturer¿s report number is 3006513822-2018-00274.
|
| |
|
Search Alerts/Recalls
|