WRIGHTS LANE SYNTHES USA PRODUCTS LLC REDUCTION FORCEPS WITH SERRATED JAW 240MM-SPEED LOCK Back to Search Results
Model Number 399.051
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A device history record review has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, one (1) reduction forceps with serrated jaw was discovered to have a broken tip on the claw.It is unknown if there was patient and procedure involvement.This report is for a reduction forceps with serrated jaw.This is report 1 of 1 for pc-(b)(4).
 
Event or Problem Description
This was discovered in sterile processing.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: event: additional information provided.Device evaluated by mfr: part: 399.051; synthes lot: t945737; release to warehouse date: may 20, 2010; manufacture site: tuttlingen; part expiration date: n/a; list of nonconformance¿s: n/a.A review of the device history records showed that there were no issues at the time of manufacturing of this device and it's sub components that would contribute to the complaint condition.The raw material certificate was reviewed, and the used material was according to the specification of the device.No nonconformance reports (ncrs) were generated during the production of this device.Review of the device history record of tuttlingen showed that there were no issues at the time of manufacturing of this device that would contribute to the issue outlined in this complaint.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDUCTION FORCEPS WITH SERRATED JAW 240MM-SPEED LOCK
Common Device Name
FORCEPS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8096722
Report Number2939274-2018-55078
Device Sequence Number2745234
Product Code HTD
UDI-Device Identifier10886982202253
UDI-Public(01)10886982202253
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number399.051
Device Catalogue Number399.051
Device Lot NumberT945737
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/02/2018
Supplement Date Received by Manufacturer11/27/2018
Initial Report FDA Received Date11/21/2018
Supplement Report FDA Received Date12/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-