| Model Number |
902814 |
| Medical Device Problem Code |
Failure to Infuse (2340)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that it was difficult to inflate the balloon during a pretest.The inflation funnel inflated.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported issue was unconfirmed.The device was returned and visually inspected.Inspected the exterior of the sample and did not find any evidence of a failure that would support the reported event.Evaluation found the catheter can be inflated without difficulty.Dissected the catheter and found no conditions that would contribute to the reported problem.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿(6) insert catheter into urethral meatus, and advance it until the balloon enters the bladder and urine flows out through the catheter.Using a syringe packaged in the tray, infuse the specified volume of sterile water into the inflation lumen to inflate the balloon.".
|
| |
|
Event or Problem Description
|
|
It was reported that it was difficult to inflate the balloon during a pretest.The inflation funnel inflated.
|
| |
|