C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER; FOLEY CATHETER (LATEX) Back to Search Results
Model Number 902814
Medical Device Problem Code Failure to Infuse (2340)
Health Effect - Clinical Codes No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Date of Event 10/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that it was difficult to inflate the balloon during a pretest.The inflation funnel inflated.
 
Additional Manufacturer Narrative
The reported issue was unconfirmed.The device was returned and visually inspected.Inspected the exterior of the sample and did not find any evidence of a failure that would support the reported event.Evaluation found the catheter can be inflated without difficulty.Dissected the catheter and found no conditions that would contribute to the reported problem.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿(6) insert catheter into urethral meatus, and advance it until the balloon enters the bladder and urine flows out through the catheter.Using a syringe packaged in the tray, infuse the specified volume of sterile water into the inflation lumen to inflate the balloon.".
 
Event or Problem Description
It was reported that it was difficult to inflate the balloon during a pretest.The inflation funnel inflated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER
Common Device Name
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8096803
Report Number1018233-2018-05602
Device Sequence Number19404580
Product Code EZC
UDI-Device Identifier00801741046100
UDI-Public(01)00801741046100
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2021
Device Model Number902814
Device Catalogue Number0165L14
Device Lot Number8ER0056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2018
Initial Date Received by Manufacturer 10/31/2018
Supplement Date Received by Manufacturer12/04/2018
Initial Report FDA Received Date11/21/2018
Supplement Report FDA Received Date12/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-