MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7015
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/08/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.A 3.25mm x 15mm quantum maverick balloon catheter was advanced for dilatation.However, the balloon ruptured.No patient complications were reported.
 
Additional Manufacturer Narrative
This supplemental correction report is being filed to correct the report number in a previously submitted supplemental report (follow-up number 001) which was accepted by fda on 12/18/2018 04:57 pm ct.The report number is being corrected from: 2134265-2018-64133 to: 2134265-2018-63102.
 
Event or Problem Description
It was reported that balloon rupture occurred.A 3.25mm x 15mm quantum maverick balloon catheter was advanced for dilatation.However, the balloon ruptured.No patient complications were reported.Device evaluated by mfr.: the device was returned for analysis.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the whole device.Microscopic examination revealed that the inflation lumen is stretched 25.1cm from the tip and is approximately 11.5cm long.The tip is damaged and the balloon was still tightly folded prior to inflation testing.There is blood present in the guidewire lumen and on the balloon folds.Review of the product specification indicates the rated burst pressure (rbp) of the complaint device.The balloon was inflated to rbp and could hold pressure for 5 minutes.Inspection of the remainder of the device presented no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8097297
Report Number2134265-2018-63102
Device Sequence Number1903007
Product Code LOX
UDI-Device Identifier08714729392798
UDI-Public08714729392798
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/11/2020
Device Model Number7015
Device Catalogue Number7015
Device Lot Number0020638037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2018
Initial Date Received by Manufacturer 10/31/2018
Supplement Date Received by Manufacturer12/06/2018
Initial Report FDA Received Date11/21/2018
Supplement Report FDA Received Date01/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-