Catalog Number 48231330
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/25/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery screwdriver was not locking properly and felt wobbly, while loading first screw.The screwdriver was not engaging properly while loading the second screw.At the end of procedure, unable to take components of one screwdriver apart.The procedure was completed successfully using back-up device.No adverse consequence to the patient was reported.
 
Additional Manufacturer Narrative
Correction: lot # corrected to 11a25.Method: visual inspection, functional inspection, product history review, complaint history review, nc/capa history review, labelling review, risk assessment result: the customer reported event of was not confirmed as the reported complaint could not be duplicated during functional testing.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.A visual inspection was performed and no visual non-conformances were identified.A test screw was loaded onto the fully assembled screwdriver.The screw was tight with no wobble noticed.The screw was driven in and out of a test block with no issues.The device functioned as expected, the reported complaint could not be duplicated.Conclusion: as the reported complaint could not be duplicated during functional testing and the device functioned as expected, a definite root cause cannot be determined.Possible root causes include incorrect alignment, screwdriver not assembled properly, or damaged screw.
 
Event or Problem Description
It was reported that during surgery screwdriver was not locking properly and felt wobbly, while loading first screw.The screwdriver was not engaging properly while loading the second screw.At the end of procedure, unable to take components of one screwdriver apart.The procedure was completed successfully using back--up device.No adverse consequence to the patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA 3 TITANIUM POLYAXIAL SCREWDRIVER (EVERYTHING BUT HANDLE)
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key8097785
Report Number0009617544-2018-00275
Device Sequence Number2307650
Product Code LXH
UDI-Device Identifier07613327265408
UDI-Public07613327265408
Combination Product (Y/N)N
Initial Reporter CountryGB
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number48231330
Device Lot Number11A255
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2018
Initial Date Received by Manufacturer 10/25/2018
Supplement Date Received by Manufacturer12/21/2018
Initial Report FDA Received Date11/21/2018
Supplement Report FDA Received Date01/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-