ENDOLOGIX AFX; BIFURCATED STENT GRAFT Back to Search Results
Model Number BA22-90/I16-30
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Codes Aneurysm (1708); Failure of Implant (1924)
Date of Event 09/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device involved in the event will not be returned for evaluation as it remains implanted in the patient.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
 
Event or Problem Description
The patient was initially implanted with a bifurcated stent graft and a suprarenal stent graft extension to treat an abdominal aortic aneurysm (aaa).This patient has been reported to have a type 2 endoleak with sac growth in the past with no intervention reported.Now it has been reported that this patient still has the type 2 endoleak with sac growth but may also have a type 3a endoleak.The physician is currently planning a an intervention.As of the date of this report, there have been no additional patient sequelae reported.
 
Additional Manufacturer Narrative
No additional investigation of this reported event is planned however if additional information pertinent to the incident is obtained, a follow-up report will be submitted.Endologix will continue to monitor this complaint and similar complaints in the event further investigation is needed.Correction: patient code: remove codes 1924 and 1708.Device code: remove code 1356.
 
Event or Problem Description
Additional information: during our investigation it was confirmed that there is no device failure therefore a complaint is not warranted since there is no alleged deficiency related to identity, quality, durability, reliability, safety, effectiveness, or performance of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key8097853
Report Number2031527-2018-00870
Device Sequence Number19336759
Product Code MIH
UDI-Device Identifier00818009010223
UDI-Public(01)00818009010223(17)180509
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/09/2018
Device Model NumberBA22-90/I16-30
Device Lot Number1341366-022
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/22/2018
Supplement Date Received by Manufacturer10/22/2018
Initial Report FDA Received Date11/21/2018
Supplement Report FDA Received Date02/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SUPRARENAL EXTENSION, LOT # 1251993008
Outcome Attributed to Adverse Event Other;
Patient Age86 YR
-
-