Model Number BEA22-60/I13-40
Medical Device Problem Code Partial Blockage (1065)
Health Effect - Clinical Codes Pulmonary Embolism (1498); Failure of Implant (1924); Occlusion (1984)
Date of Event 10/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device involved in the event will not be returned for evaluation as it remains implanted in the patient.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
 
Event or Problem Description
The patient was initially implanted with a bifurcated stent graft to treat an abdominal aortic aneurysm (aaa).Approximately two (2) years post initial implant,an afx occlusion and pulmonary embolism (pe) were detected.The physician elected to treat the patient by an aortobifem on an unknown date.As of the date of this report, there have been no additional patient sequelae reported.
 
Additional Manufacturer Narrative
The manufacturing lot evaluation confirmed all devices met specifications prior to release.The device was not returned as the device remains implanted therefore no evaluation was completed.At the completion of the clinical assessment, clinical was able to find substantial evidence to support the following reported event of right iliac to right common femoral artery occlusion.This event was most likely user related due to the off-label use.There was absence of an abdominal aneurysm, and prior endovascular repair with non-endologix products no additional investigation of this reported event is planned however if additional information pertinent to the incident is obtained, a follow-up report will be submitted.Endologix will continue to monitor this complaint and similar complaints in the event further investigation is needed.
 
Event or Problem Description
Additional information; the patient was reported as doing well and the reported event has been resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key8097859
Report Number2031527-2018-00878
Device Sequence Number2024767
Product Code MIH
UDI-Device Identifier00818009015068
UDI-Public(01)00818009015068(17)170713
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 10/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/13/2017
Device Model NumberBEA22-60/I13-40
Device Lot Number1519908-007
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/24/2018
Supplement Date Received by Manufacturer10/24/2018
Initial Report FDA Received Date11/21/2018
Supplement Report FDA Received Date02/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0007-2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
-
-