| Model Number |
45-652050 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device was discarded at the user facility and not returned to the manufacturer for review; therefore, a product analysis could not be performed.The root cause is unknown.The instructions for use (ifu) identifies premature or difficult coil detachment as potential complications associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the coil got stuck in the microcatheter and during the removal attempt, the coil detached.The coil was removed in its entirety along with the microcatheter.There was no reported patient injury or intervention.The patient was reported to be stable.
|
| |
|
Search Alerts/Recalls
|