| Catalog Number |
400SMTSTD1545 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the pusher assembly was kinked approximately 57.0 and 123.0 cm from the proximal end.The embolization coil was intact with the pusher assembly and was undamaged.Conclusions: evaluation of the returned smart coil revealed the embolization coil was detached and the distal detachment tip was sheared off from the pusher assembly.The distal detachment tip shearing from the pusher assembly was likely a result of the forceful retraction against resistance experienced during the procedure.If the distal detachment tip becomes sheared off the pusher assembly, it will result in the pull wire retracting out of the distal detachment tip, detaching the embolization coil.Further evaluation of the embolization coil revealed an unknown substance coagulated on the device in two locations.This substance may have become coagulated post-procedure within the microcatheter during return to penumbra.Evaluation of the returned smart coil revealed kinks on the pusher assembly.This damage is likely a result of forcefully advancing the device against the resistance experienced during the procedure.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2018-02297.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the external carotid artery (eca) using a penumbra smart coil (smart coil).During the procedure, while attempting to advance a smart coil through a non-penumbra microcatheter, the physician experienced resistance and decided to remove the smart coil.The physician then chose another smart coil and was able to advance it a little farther than the previous smart coil; however, resistance was encountered again and upon removal, the smart coil unintentionally detached inside the microcatheter.Therefore, the physician removed the microcatheter containing the detached coil.The procedure was completed using a new microcatheter and additional coils.There was no report of adverse effect to the patient.
|
| |
|
Search Alerts/Recalls
|