• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Catalog Number X0509160
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Levels implanted: l5/s1 it was reported that intra-op, the center t27 nut of the crosslink did not torque out and it was acted like it stripped.No health damage in the patient was reported.
 
Additional Manufacturer Narrative
Product analysis result: visual optical functional visual and optical inspection confirmed multiple witness marks along the shafts of the crossing.Functional inspection confirmed the center nut will no longer tighten the crosslink arms of the implant.It appears the center nut has been over torqued damaging the threads of the multispan eyebolt.The center nut is unable to be threaded off.This type of damage is consistent with the screw being over torqued.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key8098368
Report Number1030489-2018-01520
Device Sequence Number2325549
Product Code KWP
UDI-Device Identifier00613994648914
UDI-Public00613994648914
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K030840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberX0509160
Device Lot Number0565718W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2019
Initial Date Received by Manufacturer 10/25/2018
Supplement Date Received by Manufacturer01/11/2019
Initial Report FDA Received Date11/22/2018
Supplement Report FDA Received Date02/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-