• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ENDURANT BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ENDURANT BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number ENBF3620C145E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Injury (2348)
Date of Event 09/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of medical devices: information references the main component of the system.Other relevant device(s) are: product id: enew2020c82e, serial/lot #: (b)(4), ubd: 16-aug-2013, udi#: (b)(4); product id: enlw1620c124e, serial/lot #: (b)(4), ubd: 30-mar-2013, udi#: (b)(4).Month and year valid.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant stent graft system was implanted in a patient for the endovascular treatment of a 69mm diameter abdominal aortic aneurysm.It was reported that a ct scan carried out approximately six years later revealed an unknown endoleak.An endurant cuff and two endurant limbs were inserted but the endoleak was visible above the graft but not visible when inside the graft.The physician elected to implant endoanchors and a non mdt stent graft however the endoleak has not resolved and the physician ended the procedure at this time.Per the physician the cause of the event was possibly due to disease progression.No additional clinical sequelae were reported and the patient will be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT BIFURCATED STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key8098470
Report Number2953200-2018-01600
Device Sequence Number13931610
Product Code MIH
UDI-Device Identifier00613994762368
UDI-Public00613994762368
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/05/2013
Device Model NumberENBF3620C145E
Device Catalogue NumberENBF3620C145E
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2018
Initial Report FDA Received Date11/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
-
-