• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT150
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Blurred Vision (2137); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2018 and (b)(6) 2018.(b)(4).Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the placement of the zxt150 intraocular lens (iol) in the left eye of the patient was noticed to be misaligned during the post-op exam.The patient complained of blurry vision and the iol was explanted on (b)(6) 2018.The incision was enlarged to remove the lens and sutures were used; no vitrectomy was performed.The patient was reported being in good condition following the surgery.The account commented that the explanted lens was discarded.Additional information was received and it was learnt that the lens was initially planned to be repositioned.However, the surgeon then determined that the patient's eye would not hold the lens and decided to explant the lens instead.No additional information provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-(b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key8098522
Report Number9614546-2018-01110
Device Sequence Number14034848
Product Code POE
UDI-Device Identifier05050474605282
UDI-Public(01)05050474605282(17)220523
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/05/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/23/2022
Device Model NumberZXT150
Device Catalogue NumberZXT150U240
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/30/2018
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date11/22/2018
Supplement Report FDA Received Date11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-