| Model Number |
MMT-523LNAB |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement test.Unit received with motor error alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform prime/a33 test, excessive no delivery test and occlusion test due to motor error alarm.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had motor error.The customer¿s blood glucose level was 150 mg/dl at the time of the incident.Customer was able to successfully clear the alarm.Customer was able to complete insulin pump rewind.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Search Alerts/Recalls
|