Batch review performed on (b)(6) 2018.Lot 168448: (b)(4) items manufactured and released on 07 march 2017.Expiration date: 02/23/2022.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Additional component revised: liner: mectacer 01.29.414 ceramic liner ø 36 / f (not marketed in us), lot 157799: (b)(4) items manufactured and released on 04 april 2016.Expiration date: 2021-03-13.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Ball heads: mectacer 01.29.209 biolox delta ceramic ball head 12/14 ø 36 size m 0 , (k112115) lot 177707: (b)(4) items manufactured and released on 17 april 2017.Expiration date: 2023-03-27.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
|
|
Two months and a half after primary, while the patient was tying the shoe, the hip dislocated (liner/head from cup).The hip was reducted but then dislocated a second time, so the surgeon revised the implant.The surgeon was advised by another orthopedic surgeon who he works with regularly to use a more constrained option for the cup; the second surgeon assisted him during the revision surgery and opted to used zimmer cup (which is what he was used to using).
|