MAUDE Adverse Event Report: COOK IRELAND LTD RESONANCE STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD RESONANCE STENT SET; FAD STENT, URETERAL Back to Search Results
Model Number G34109
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/25/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
510 (k) number: k063742.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
Rms- stent exchange was performed on both right and left sides after 12 months of placement.One stent could be removed with no problem, then exchanged.The physician attempted to remove another stent as well after grasping the stent with forceps, but he felt resistance during the attempt of stent removal.He managed to remove the stent, but it broke at the renal side and the separated segment remained inside the patient.No treatment was conducted at physician's judgment and a new rms- stent was placed to complete the procedure.There have been no adverse effects to the patient reported.Updated on (b)(6) 2018, on (b)(6) 2017 (exact date unknown): rms-060022-r/ lot# c1296687 (complaint device) and rms-060022-r/ lot# c1307377 were placed in the right and left sides respectively.There had been no problem observed in the renal function and devices performance of drainage during placement.On (b)(6) 2018: rms- stent exchange was performed on both right and left sides.One stent (c1307377) could be removed with no problem, then exchanged.The physician attempted to remove another stent (c1296687) as well after grasping the stent with forceps, but he felt resistance during the attempt of stent removal.He managed to remove the stent, but it broke at the renal side and the separated segment remained inside the renal pelvis.No treatment was conducted at physician's judgment and a new rms- stent was placed to complete the procedure.As of (b)(6) 2018: the patient had a favorable outcome without bleeding after stents exchange.The physician has a plan to retrieve the remained segment from the renal pelvis percutaneously through the nephrostomy route since the pre existing disease is retroperitoneal fibrosis which is benign and good prognosis can be expected.However, the patient does not particularly require extraction of the remained segment since no problem has occurred.There have been no adverse effects to the patient reported.
 
Event or Problem Description
Rms- stent exchange was performed on both right and left sides after 12 months of placement.One stent could be removed with no problem, then exchanged.The physician attempted to remove another stent as well after grasping the stent with forceps, but he felt resistance during the attempt of stent removal.He managed to remove the stent, but it broke at the renal side and the separated segment remained inside the patient.No treatment was conducted at physician's judgment and a new rms- stent was placed to complete the procedure.There have been no adverse effects to the patient reported.Updated on 30/oct/2018, on (b)(6) 2017 (exact date unknown): rms-060022-r/ lot#: c1296687 (complaint device) and rms-060022-r/ lot#: c1307377 were placed in the right and left sides respectively.There had been no problem observed in the renal function and devices performance of drainage during placement.On (b)(6) 2018: rms- stent exchange was performed on both right and left sides.One stent (c1307377) could be removed with no problem, then exchanged.The physician attempted to remove another stent (c1296687) as well after grasping the stent with forceps, but he felt resistance during the attempt of stent removal.He managed to remove the stent, but it broke at the renal side and the separated segment remained inside the renal pelvis.No treatment was conducted at physician's judgment and a new rms- stent was placed to complete the procedure.As of on (b)(6) 2018: the patient had a favorable outcome without bleeding after stents exchange.The physician has a plan to retrieve the remained segment from the renal pelvis percutaneously through the nephrostomy route since the pre existing disease is retroperitoneal fibrosis which is benign and good prognosis can be expected.However, the patient does not particularly require extraction of the remained segment since no problem has occurred.There have been no adverse effects to the patient reported.Additional information provided on (b)(6) 2018, (b)(6).It was confirmed with the physician that the remained segment would be retrieved from the patient via the percutaneous approach on (b)(6) 2019.The segment will be returned to cook for investigation.Additional information (25/jan/2019, (b)(6).On (b)(6) 2019: the rms- stent which had been placed on (b)(6) 2018 after the reported event was removed transurethrally with no problem.On (b)(6) 2019: the physician performed retrieval of the remained segment as planned.The patient conditions including renal function had been excellent.A nephrostomy tract was established with an ultraxx nephrostomy balloon percutaneously.Though the tip of the remained segment looked as if it had been embedded in the distal renal calix under fluoroscopic guidance prior to establishing the nephrostomy tract, pyeloscope showed the segment right in front, which made the physician impossible to observe how the tip had been positioned.The segment was retrieved with forceps successfully and found to be so clean as it looked like an unused stent.A nephrostomy balloon was placed after that, then the procedure was finished.The placed nephrostomy balloon is scheduled to be removed in one or two days and the flow of urine is checked in order to decide which stent, rms- or other models, should be placed next or whether wait-and-see approach would be preferable.There have been no adverse effects to the patient reported.
 
Additional Manufacturer Narrative
510 (k) number: k063742.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4).Cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.The device was returned pending evaluation, an additional mdr will be submitted to include the evaluation findings.
 
Event or Problem Description
Rms- stent exchange was performed on both right and left sides after 12 months of placement.One stent could be removed with no problem, then exchanged.The physician attempted to remove another stent as well after grasping the stent with forceps, but he felt resistance during the attempt of stent removal.He managed to remove the stent, but it broke at the renal side and the separated segment remained inside the patient.No treatment was conducted at physician's judgment and a new rms- stent was placed to complete the procedure.There have been no adverse effects to the patient reported.Updated on 30/oct/2018, (b)(6) 2017 (exact date unknown): rms-060022-r/ lot# c1296687 (complaint device) and rms-060022-r/ lot# c1307377 were placed in the right and left sides respectively.There had been no problem observed in the renal function and devices performance of drainage during placement.(b)(6) 2018: rms- stent exchange was performed on both right and left sides.One stent (c1307377) could be removed with no problem, then exchanged.The physician attempted to remove another stent (c1296687) as well after grasping the stent with forceps, but he felt resistance during the attempt of stent removal.He managed to remove the stent, but it broke at the renal side and the separated segment remained inside the renal pelvis.No treatment was conducted at physician's judgment and a new rms- stent was placed to complete the procedure.As of (b)(6) 2018: the patient had a favorable outcome without bleeding after stents exchange.The physician has a plan to retrieve the remained segment from the renal pelvis percutaneously through the nephrostomy route since the pre existing disease is retroperitoneal fibrosis which is benign and good prognosis can be expected.However, the patient does not particularly require extraction of the remained segment since no problem has occurred.There have been no adverse effects to the patient reported.
 
Additional Manufacturer Narrative
510 (k) number: k063742.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4).Cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.1 x rms-060022-r of lot number c1296687 was not returned to cirl for an evaluation therefore a limited documentation based investigation was completed.An image review was completed on 20th of november, the summary from the image review is a fracture of a resonance metallic ureteral stent during attempted exchange after a dwell time exceeding the ifu recommended dwell time.The right stent fractured near the renal pelvis when resistance was felt by the physician.The distal component remained inside the patient, given the point of fracture, this likely indicates there was either stent encrustation or tissue ingrowth involving the renal portion of the stent, inhibiting its removal.There was no discussion in the complaint report regarding any clinical scenarios that may have facilitated either these 2 scenarios and is therefore likely related to the prolonged dwell time.A possible cause of this complaint could be attributed to the dwell time which exceeded the recommended dwell time of 12 months and it is possible there was either stent encrustation or tissue ingrowth involving the renal portion of the stent, inhibiting its removal as a result of this extended dwell time.However, as the device was not returned and the clinical conditions cannot be replicated it is not possible to conclusively determine the root cause of this complaint.A review of the manufacturing records for rms-060022-r of lot # c1296687 did not reveal any issues associated with this complaint.There is no evidence to suggest that this issue affects the entire lot # c1296687; upon review of complaints this failure mode has not occurred previously with this lot # c1296687.The instructions for use, ifu0020-15 which accompanies this device instructs the user ¿ the stent must not remain indwelling more than twelve months.If the patient¿s status permits, the stent may be replaced with a new stent.¿ also it warns ¿individual variations of interactions between stents and the urinary system are unpredictable.Patients should be checked at regular intervals utilizing techniques such as abdominal x-ray (kub film).Patient using calcium supplements must be more closely monitored for possible stent encrustation.The stent must be removed if encrustation hampers drainage.¿ also the instruction for use warns the user ¿if any resistance is encountered when removing the ureteral stent, determine the cause of the resistance by such as x-ray examination and perform appropriate treatment.Forceful removal may result in damage to renal pelvis and the ureter.¿ the stents were placed in (b)(6) 2017 and removal was (b)(6) 2018, which exceeds the indwell time as outlined in the instructions for use which accompanies this device.The user did not follow the instruction for use for this device.A notification of user error has been send to the product manager see attached email customer complaint is confirmed based on the customer testimony.The patient didn¿t experience any adverse as result of the device failure and based on the information provided.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Additional Manufacturer Narrative
510 (k) number: k063742.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031 information pertaining (b)(4).Importer site establishment registration number: (b)(4).1.Device evaluation: 1 x rms-060022-r of lot number c1296687 was returned to cirl for an evaluation.It was returned used and broken and it was not returned in its original packaging.2.Lab evaluation: 1 x rms-060022-r of lot number c1296687 was returned to cirl for an evaluation after some time after the original investigation was concluded.The device involved in the complaint was evaluated in laboratory on (b)(6) 2019.The returned device lab examination findings and observations can be referred through attached photos.In summary the following results were observed in the lab evaluation.A broken part of the stent, it was approx.5cm in length was returned the complaint is confirmed as the failure was verified in the laboratory and through previously submitted the images.3.Image review: an image review was completed on (b)(6) see attached ¿(b)(4) imaging review ver1.Pdf¿.The summary from the image review is a fracture of a resonance metallic ureteral stent during attempted exchange after a dwell time exceeding the ifu recommended dwell time.The right stent fractured near the renal pelvis when resistance was felt by the physician.The distal component remained inside the patient, given the point of fracture, this likely indicates there was either stent encrustation or tissue ingrowth involving the renal portion of the stent, inhibiting its removal.There was no discussion in the complaint report regarding any clinical scenarios that may have facilitated either these 2 scenarios and is therefore likely related to the prolonged dwell time.4.Root cause (possible): a possible cause of this complaint could be attributed to the dwell time which exceeded the recommended dwell time of 12 months and it is possible there was either stent encrustation or tissue ingrowth involving the renal portion of the stent, inhibiting its removal as a result of this extended dwell time.5.Documents review: a review of the manufacturing records for rms-060022-r of lot # c1296687 did not reveal any issues associated with this complaint.There is no evidence to suggest that this issue affects the entire lot # c1296687; upon review of complaints this failure mode has not occurred previously with this lot # c1296687.These inspections and functional checks are outlined in internal procedures in place at cirl step 1.1 ¿check for kinks or damage along stent length,¿ step 1.4 ¿ensure smooth and clean (non- discolored) welded ends¿ and step 1.5 ¿tug at both ends of the stent to confirm the inner wire to coiled wire connection.¿ the instructions for use, ifu0020-15 which accompanies this device instructs the user ¿ the stent must not remain indwelling more than twelve (12 months.If the patient¿s status permits, the stent may be replaced with a new stent.¿ also it warns ¿individual variations of interactions between stents and the urinary system are unpredictable.Patients should be checked at regular intervals utilizing techniques such as abdominal x-ray (kub film).Patient using calcium supplements must be more closely monitored for possible stent encrustation.The stent must be removed if encrustation hampers drainage.¿ also the instruction for use warns the user ¿if any resistance is encountered when removing the ureteral stent, determine the cause of the resistance by such as x-ray examination and perform appropriate treatment.Forceful removal may result in damage to renal pelvis and the ureter.¿ the stents were placed in (b)(6) 2017 and removal was (b)(6) 2018, which exceeds the indwell time as outlined in the instructions for use which accompanies this device.The user did not follow the instruction for use for this device.6.Summary: complaint is confirmed as the failure was verified in the laboratory.The patient didn¿t experience any adverse as result of the device failure and based on the information provided.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
Rms- stent exchange was performed on both right and left sides after 12 months of placement.One stent could be removed with no problem, then exchanged.The physician attempted to remove another stent as well after grasping the stent with forceps, but he felt resistance during the attempt of stent removal.He managed to remove the stent, but it broke at the renal side and the separated segment remained inside the patient.No treatment was conducted at physician's judgment and a new rms- stent was placed to complete the procedure.There have been no adverse effects to the patient reported. it was confirmed with the physician that the remained segment would be retrieved from the patient via the percutaneous approach on (b)(6) 2019.The segment will be returned to cook for investigation. (b)(6) 2019: the rms- stent which had been placed on (b)(6) 2018 after the reported event was removed transurethrally with no problem.(b)(6) 2019: the physician performed retrieval of the remained segment as planned.The patient conditions including renal function had been excellent.A nephrostomy tract was established with an ultraxx nephrostomy balloon percutaneously.Though the tip of the remained segment looked as if it had been embedded in the distal renal calix under fluoroscopic guidance prior to establishing the nephrostomy tract, pyeloscope showed the segment right in front, which made the physician impossible to observe how the tip had been positioned.The segment was retrieved with forceps successfully and found to be so clean as it looked like an unused stent.A nephrostomy balloon was placed after that, then the procedure was finished.The placed nephrostomy balloon is scheduled to be removed in one or two days and the flow of urine is checked in order to decide which stent, rms- or other models, should be placed next or whether wait-and-see approach would be preferable.There have been no adverse effects to the patient reported.Fda mdr reporting required - reporting required based on the requirement for a surgical intervention as the stent broke on removal from the patient and the physician "plans to retrieve the remained segment from the renal pelvis percutaneously through the nephrostomy route" no adverse effects to the patient have been reported as occurring.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONANCE STENT SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key8098868
Report Number3001845648-2018-00539
Device Sequence Number2364555
Product Code FAD
UDI-Device Identifier10827002341095
UDI-Public(01)10827002341095(17)191121(10)C1296687
Combination Product (Y/N)N
PMA/510(K) Number
K063742
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/21/2019
Device Model NumberG34109
Device Catalogue NumberRMS-060022-R
Device Lot NumberC1296687
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date11/22/2018
Event Location Hospital
Initial Date Received by Manufacturer 10/26/2018
Supplement Date Received by Manufacturer10/26/2018
10/26/2018
10/26/2018
Initial Report FDA Received Date11/22/2018
Supplement Report FDA Received Date12/20/2018
02/21/2019
03/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-