| Catalog Number |
RONYX20008UX |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An attempt was made to use a resolute onyx rx coronary drug eluting stent to treat a moderately tortuous and calcified lesion exhibiting 90% stenosis located in the ostium of the saphenous vein graft (svg).The device was inspected with no issues.Negative prep was not performed.The lesion was pre-dilated.The device passed through a previously deployed stent.Resistance was encountered when advancing the device.Excessive force was used during delivery.It was reported that stent dislodgement occurred during removal following a failed delivery.The dislodged stent was not removed from the patient.There was no intervention attempted to remove the dislodged stent.The patient is reported to be alive with no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
There was no damage to the previously deployed stent.If information is provided in the future, a supplemental report will be issued.
|
| |
|