Visual assessment under magnification confirmed the plating and silicone that is applied appeared normal.The inner blade exhibited an abrasion at the tip along with a corresponding abrasion on the inner blade.The inner edge form appeared to have some damage to the cutting edge surface; this damage may have resulted from making contact with another object.The damage on the inner edgeform radius seems to have created an uneven surface that yielded some excess material that could be visually seen protruding above the cutting surface.The excess material appears to have made contact with the id of the outer edgeform which potentially created some galling/shedding between the two surfaces to occur.There was no friction felt during the functional spin test while the inner blade rotated freely within the outer blade in the unloaded condition.Per the evaluation that was performed based on the details that were provided, the investigation has confirmed that material has been displaced between the two surfaces (inner/outer edgeform) as described by the customer resulting in some material galling/shedding to occur.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: excessive ¿side loading¿ on the blade during use does not improve cutting performance and in extreme cases may result in wear and degradation of the inner assembly.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.A review of relevant clinical/medical information in the reported issue, inclusive of technique and patient information, to include, but not limited to: patient information surgical procedure/post-operative care review device labeling (including technique guides, ifus, etc.) it was reported that the small metal chips from the blade were removed upon flushing the joint, and no patient harm or injury was sustained as a result.Based on the information provided, no additional risk to the patient is anticipated.No further medical assessment is warranted based on the information provided.
|