| Model Number |
459801419351 |
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The wedge id is not exported via dicom unless manufacture code is populated in physics (300a,00d4).Initial investigation resulted in the realization that there was no manufacture code for the edw wedge in physics.This resulted in no wedge id being exported via dicom to the r&v system.
|
| |
|
Additional Manufacturer Narrative
|
|
Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
| |
|