SMITH & NEPHEW, INC. JOURNEY TIBIA BASE NP RT SZ 5; PRSTHSS, KNEE, PTLLFMRTBIAL, SM-CNSTRND, CMNTD, POLYMER/METAL/POLYMER Back to Search Results
Catalog Number 74022215
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 10/24/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that revision surgery was performed due to insert disassociation from baseplate.During revision surgery, the surgeon found that bone cement intervened between baseplate and insert.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.The clinical/medical team concluded, it has been communicated that no clinical documents are available.Therefore, no thorough clinical assessment of the reported issue can be rendered.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed parts revealed no prior complaints for the listed batches.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY TIBIA BASE NP RT SZ 5
Common Device Name
PRSTHSS, KNEE, PTLLFMRTBIAL, SM-CNSTRND, CMNTD, POLYMER/METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8100388
Report Number1020279-2018-02572
Device Sequence Number2389475
Product Code JWH
UDI-Device Identifier03596010554291
UDI-Public03596010554291
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K042515
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number74022215
Device Lot Number17FM17356
Initial Date Received by Manufacturer 10/30/2018
Supplement Date Received by Manufacturer10/30/2018
Initial Report FDA Received Date11/25/2018
Supplement Report FDA Received Date01/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
74022116 / 17HM1551771420576 / 17DM07125
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-