• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTSINE TECHNOLOGIES LTD PACKAGE,450P,PP01,EN,450-BAC-US-10; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTSINE TECHNOLOGIES LTD PACKAGE,450P,PP01,EN,450-BAC-US-10; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Catalog Number 80515-000002
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2015058.Heartsine technologies ltd (manufacturer) is submitting the report on behalf of heartsine technologies llc (importer).
 
Event or Problem Description
Device service required prompt.There was no patient involved in this event.
 
Additional Manufacturer Narrative
The device history records for the sam 450p device and pad-pak were reviewed and this confirmed that all manufacturing and quality checks and test had been successfully completed.The sam 450p passed ¿out qat from heartsine technologies on the (b)(6)2018.A fractured joint on j12 pin 14 (key 3) of the membrane pba resulted in the device failing self-tests due to the device not powering up in 450p cpr advisor mode.These self-test fails occurred between the (b)(6)2018 and the(b)(6)2018.The user would have been alerted with a ¿warning, device service required¿ prompt and a flashing red status led, as per the reported fault.The device successfully passed final unit test, h041-014-004, on the (b)(6)2018.The device successfully passed all weekly auto self-tests from the(b)(6)2018 up to the (b)(6)2018.Therefore, it is assumed the solder joint was making sufficient contact during these times.It is the policy of heartsine not to refurbish devices which have been returned from the field, after investigation, therefore this device shall be scrapped and replaced with another device of the same model.
 
Event or Problem Description
Device service required prompt.There was no patient involved in this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGE,450P,PP01,EN,450-BAC-US-10
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
HEARTSINE TECHNOLOGIES LTD
203 airport road west
belfast BT3 9 ED
MDR Report Key8100415
Report Number3004123209-2018-00709
Device Sequence Number13789579
Product Code MKJ
UDI-Device IdentifierM727SAM450P
UDI-PublicM727SAM450P
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P160008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number80515-000002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2019
Initial Date Received by Manufacturer 11/09/2018
Supplement Date Received by Manufacturer07/11/2019
Initial Report FDA Received Date11/25/2018
Supplement Report FDA Received Date07/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-