| Catalog Number |
859-545 |
| Medical Device Problem Code |
Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was pre-operatively diagnosed with vertebral fracture and underwent lumbar internal fixation surgery.During surgery, the screw slipped.The product came in contact with the patient.No fragment remains in the patient.No patient symptoms or complication were reported as result of this event.The patient has achieved solid fusion.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis result: visual optical functional visual and optical examination revealed thread crest/flank damage; this damage appears to have initiated near the start of the thread, and is evident around the damaged portion of the thread.Functional evaluation with a sample set screw confirmed the fas screw was unable to be fully engaged.The above observations are consistent with misalignment of the fas and set screw threads during construct assembly.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|