| Model Number |
MMT-751PNAL |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Adverse Event Without Identified Device or Use Problem (2993); Power Problem (3010); Priming Problem (4040)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had keypad anomaly prime anomaly.The customer¿s blood glucose level was unknown.The customer reported that the insulin pump had rejected a brand new battery before during priming and it was leaked and drips more than it should.The customer reported that the buttons were not as responsive; harder to see screen.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with no (act, up and escape) button response due to corroded keypad traces.No button error alarm during testing noted.Unable to perform all functional testing including the displacement, operating currents, a21 error test, rewind, basic occlusion, occlusion, prime a33 and excessive no delivery test or verify prime/fill anomaly and failed battery test due to buttons not responding.(b)(4).
|
| |
|