Model Number MMT-522LNAL
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/21/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pump had unresponsive buttons at esc and act due to unlocked j2/liquid crystal display keypad connector during visual inspection.Unable to perform operating currents, self test, unexpected restart alarm error test, rewind test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test or verify operating currents and motor position encoder error alarm due to unresponsive button.No button error alarm during testing.However, corrosion was found at the keypad traces.The motor was tested outside of the device and passed.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had motor error alarm.The blood glucose level was 160 mg/dl.The insulin pump will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522LNAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key8101369
Report Number2032227-2018-66502
Device Sequence Number11099204
Product Code OYC
UDI-Device Identifier00613994407719
UDI-Public(01)00613994407719
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-522LNAL
Device Catalogue NumberMMT-522LNAL
Device Lot NumberA000113848
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2018
Initial Report FDA Received Date11/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Patient Weight78
-
-